The ELB series high-precision pressure gauge calibration bench utilizes a liquid medium (such as water or oil) and is designed for the calibration and testing of instruments including high-precision dial pressure gauges (Bourdon tube type), pressure sensors, pressure transmitters, pressure switches, and digital manometers. It offers a control accuracy of ±0.2% of full scale; specifically, the error remains within 0.2% of the gauge’s upper range limit during both the ascending and descending pressure phases of calibration, while consistent repeatability is maintained throughout the entire process.
ELB Series High Precision Pressure Gauge Calibration Bench Selection Table
| Model | Applicable Pressure Gauge Range (MPa) | Standard Volume | Minimum Set Pressure | Maximum Pressure | Control Mode (Multiple Selection) | ||
|---|---|---|---|---|---|---|---|
| Manual / Automatic Pressure Regulation | Multi-point Selection | Multi-point Cycle | |||||
| ELB-45M | 2.5, 4, 6, 10, 16, 25 | 40ml | 0.25MPa | 45MPa | Optional | Servo Valve | Servo Valve |
| ELB-90M | 6, 10, 16, 25, 40, 60 | 40ml | 0.6MPa | 90MPa | Optional | Servo Valve | Servo Valve |
| ELB-135M | 10, 16, 25, 40, 60, 100 | 40ml | 1MPa | 135MPa | Optional | Servo Valve | Servo Valve |
| ELB-210M | 16, 25, 40, 60, 100, 160 | 40ml | 1.6MPa | 210MPa | Optional | Servo Valve | Servo Valve |
Applications
- High-precision pressure gauge calibration
- Pressure sensor calibration
- Pressure transmitter calibration
- Pressure switch testing
- Digital pressure gauge calibration
- Pressure instrument verification
Key Advantages
- High Precision – Control accuracy of up to ±0.2% of full scale.
- Liquid Medium – Uses water, oil, or other suitable liquid media.
- Stable Pressure Control – Maintains consistent pressure during both ascending and descending calibration.
- Excellent Repeatability – Provides reliable and repeatable calibration results.
- Wide Application – Suitable for gauges, sensors, transmitters, switches, and digital manometers.
- High-Precision Calibration – Suitable for applications requiring accurate and reliable pressure calibration.